Decay Time
The charge-sensitive preamplifier of an HPGe detector converts the collected charge into a voltage step that relaxes back to the baseline with an exponential tail,
\[V(t) = V_0 + A \, e^{-(t - t_s)/\tau} \quad (t > t_s),\]
with the decay time $\tau$ of the preamplifier. Pole-zero (PZ) correction cancels this tail so that the waveform after the rise stays flat. All energy filters of the HPGe chains (HPGe DSP Chains) are applied to the PZ-corrected waveform, so they need a good value of $\tau$. The package determines $\tau$ per waveform from the slope of the logarithm of the tail, and the production dataflow combines the values of many waveforms into one decay time per detector.
Tail Statistics
tailstats fits a straight line to the natural logarithm of the samples in a time window $[t_1, t_2]$:
\[\ln y(t) = a + b\,t, \qquad \tau = -1/b.\]
It returns a named tuple with three fields:
| Field | Meaning |
|---|---|
mean | Mean of $\ln y$ in the window |
sigma | Standard deviation of $\ln y$ in the window |
τ | Decay time $-1/b$ in the unit of the time axis |
The window must lie in the tail, after the rise has finished and before the waveform ends. The waveform must be baseline-subtracted, because the logarithm is taken of the sample values themselves. If any sample in the window is zero or negative, the logarithm is undefined and tailstats returns zero for all three fields. The estimate is biased when the noise is not small compared to the signal in the window, because the logarithm of a noisy sample is not symmetric around the logarithm of the mean; waveforms of large amplitude give the best estimates.
using LegendDSP
using RadiationDetectorSignals, RadiationDetectorDSP
using Unitful, StableRNGs, Statistics, ArraysOfArrays, IntervalSets
rng = StableRNG(7)
τ_true = 450.0u"µs"
function synthetic_waveform(amplitude; n = 8192, dt = 16.0u"ns", t_rise = 50.0u"µs", rise = 1.0u"µs", σ = 3.0)
t = range(0.0u"ns", step = dt, length = n)
x = ustrip.(u"µs", t) .- ustrip(u"µs", t_rise)
shape = map(x) do xi
xi < 0 ? 0.0 : xi < ustrip(u"µs", rise) ? xi / ustrip(u"µs", rise) :
exp(-(xi - ustrip(u"µs", rise)) / ustrip(u"µs", τ_true))
end
RDWaveform(t, 1000 .+ amplitude .* shape .+ σ .* randn(rng, n))
end
wvfs = ArrayOfRDWaveforms([synthetic_waveform(4000 + 50 * i) for i in 1:50])
bl_window = 0.0u"µs" .. 39.0u"µs"
tail_window = 70.0u"µs" .. 110.0u"µs"
# Baseline-subtracted first waveform
wvf = shift_waveform(wvfs[1], -signalstats(wvfs[1], leftendpoint(bl_window), rightendpoint(bl_window)).mean)
stats = tailstats(wvf, leftendpoint(tail_window), rightendpoint(tail_window))
(τ = uconvert(u"µs", stats.τ), mean_of_log = stats.mean, sigma_of_log = stats.sigma)(τ = 449.3515034067366 μs, mean_of_log = 8.219806747326931, sigma_of_log = 0.02572019585422156)Decay Times of a Waveform Set
dsp_decay_times does the baseline subtraction and the tail fit for a set of waveforms. It has two methods:
dsp_decay_times(wvfs, bl_window, tail_window)
dsp_decay_times(wvfs, config::DSPConfig)The second method takes the two windows from config.bl_window and config.tail_window. Both windows are closed intervals of Unitful times. The function decodes encoded waveforms, subtracts the mean of the samples in bl_window from each waveform, calls tailstats in tail_window, and returns the decay times as a vector of quantities in µs, one per waveform. Waveforms for which tailstats returns zero (non-positive samples in the window) give 0.0 µs.
taus = dsp_decay_times(wvfs, bl_window, tail_window)
(median = median(taus), minimum = minimum(taus), maximum = maximum(taus))(median = 449.97775320115784 μs, minimum = 449.3515034067366 μs, maximum = 450.59792576478327 μs)The distribution of decay times of a set of waveforms is narrow and well described by a single peak. The scatter is driven by noise and by the amplitude of the waveforms. The production dataflow selects waveforms from a high-energy gamma line after a basic quality selection (see dsp_qc_flt_optimization_compressed and Use in Juleana.jl), computes dsp_decay_times on the presummed waveforms, restricts the values to a plausible interval, fits a truncated Gaussian to their histogram, and stores the center of the fit as the decay time of the detector.
The decay time is independent of the amplitude scale. For compressed raw data (see Waveforms and Configuration) the presummed waveform is passed to dsp_decay_times: the baseline subtraction and the logarithmic slope are the same as for the full-rate waveform, only the time axis step differs.
Use in the Pole-Zero Correction
The decay time enters all chains as the argument τ; dsp_icpc and dsp_icpc_compressed build InvCRFilter(τ) from it and apply it to the baseline-subtracted waveforms before any energy filter. The same holds for the filter optimization scans (Filter Optimization).
InvCRFilter(τ) is the inverse of a CR high-pass filter with time constant $\tau$. With the correct $\tau$ the exponential tail turns into a constant, so that the slope of the waveform in the tail window vanishes. With a $\tau$ that is too large the correction is too weak and the tail keeps a negative slope; with a $\tau$ that is too small the corrected waveform rises in the tail. The chains write the mean, standard deviation and slope of the corrected tail window to tailmean, tailsigma and tailslope so that the quality of the correction can be checked for every event.
τ_est = median(taus)
function tail_slope(τ)
corrected = InvCRFilter(τ).(shift_waveform.(wvfs, -signalstats.(wvfs, leftendpoint(bl_window), rightendpoint(bl_window)).mean))
median(signalstats.(corrected, leftendpoint(tail_window), rightendpoint(tail_window)).slope)
end
[(τ = τ, slope = tail_slope(τ)) for τ in (0.9 * τ_est, τ_est, 1.1 * τ_est)]3-element Vector{@NamedTuple{τ::Unitful.Quantity{Float64, 𝐓, Unitful.FreeUnits{(μs,), 𝐓, nothing}}, slope::Unitful.Quantity{Float64, 𝐓^-1, Unitful.FreeUnits{(ns^-1,), 𝐓^-1, nothing}}}}:
(τ = 404.9799778810421 μs, slope = 0.0011955825440521936 ns^-1)
(τ = 449.97775320115784 μs, slope = -5.758917262015172e-9 ns^-1)